Views, angles, sample volumes, and the small things that change your number.
Pulsed-wave Doppler gives you a mechanical proxy for the electrical PR interval. Understanding the gap between the two changes how you place your cursor.
What the ECG measures. Onset of atrial depolarisation to onset of ventricular depolarisation.
Not available in routine fetal practice
What Doppler gives you. Consistently longer than the true electrical PR by approximately 14–16 ms — a systematic, predictable difference, not random error.
Mechanical proxy — validated across multiple cohorts
Figure 1. Relationship between true electrical PR (ECG) and mPR surrogate. The mPR consistently exceeds the electrical PR by approximately 14–16 ms. This is a systematic, predictable difference and must be kept in mind when interpreting values against clinical thresholds.
Figure 2. At a clinical treatment threshold of 150 ms, a 10–15 ms acquisition error is not a technical footnote. It dictates treatment.
The most widely used technique with the largest normative dataset. Your starting point in every study.
Figure 3. Sample volume positioning at the anterior mitral leaflet hinge point within the LVOT. GA 25w2d. The gate must simultaneously capture mitral inflow and aortic outflow — not deeper, not at the leaflet tips.
Figure 4. LV inflow/outflow Doppler trace. A-wave above baseline (atrial contraction), aortic ejection below. PR Interval 109 ms. GA 25w2d, 150 mm/s.
Sweep slightly cephalad from 4-chamber until the aortic root appears. Fetal position must be apex-up or apex-down. A lateral heart axis makes reliable acquisition impossible — wait or ask the mother to reposition.
This is the minimum required to visualise valve clicks and time waveform onsets with precision. Below 100 mm/s, onset timing is guesswork.
Position at the anterior mitral leaflet hinge point within the LVOT. Not deeper (A-wave disappears), and not at the leaflet tips (outflow signal lost). Both A-wave and aortic ejection signal must be clearly visible.
At heart rates above approximately 160 bpm, E and A waves fuse and the A-wave onset becomes unidentifiable. Stop and switch to SVC/Ao.
Cursor 1 at the foot of the A-wave. Cursor 2 at the leading edge of the aortic opening click. Average 3 consecutive artefact-free beats. Single-beat measurement is not acceptable.
The reference range you compare against is FHR and GA dependent. Document both alongside every mPR value. Never mix methods across serial studies in the same patient.
Start of E-wave. Not used for PR measurement.
Follows atrial contraction. Best ventricular landmark — marks IVC onset. Correlation r=0.92 vs ECG (Bergman 2006). No fetal normative dataset yet.
Start of ejection. Leading edge of click base is your landmark. Click base spans ~8–15 ms at 150 mm/s — choose consistently and document it.
Not used for PR measurement.
The go-to when E-A fusion occurs or LV in/out alignment is limited. Requires near-zero insonation angle and wall filter at minimum.
Figure 5. Colour Doppler showing SVC and ascending aorta (Ao) with IVC visible. GA 22w5d. The SVC runs parallel to the Ao over a short mediastinal segment, enabling simultaneous sampling.
Figure 6. SVC/Ao Doppler trace. SVC a-wave reversal (A, above baseline) marks atrial systole. Aortic ejection (Ao, below baseline) marks ventricular ejection. GA 22w5d.
From the 3-vessel view, sweep slightly caudad, or use a near-sagittal approach when the fetus is spine-posterior. SVC runs to the right of and slightly posterior to the ascending aorta. Both must run parallel to the beam.
This is the most common reason the SVC a-wave is absent. The default wall filter will suppress it. Drop to the lowest available setting before placing the gate. If the a-wave is absent, check filter first.
Use colour Doppler to confirm both SVC and aorta are within the gate. SVC flow (venous reversal) is above baseline; aortic ejection is below.
A small brief reversal in SVC flow during atrial systole — above baseline and short in duration. It appears just before the aortic ejection waveform. If absent, filter is too high or angle is off.
Onset of SVC reversal to onset of aortic ejection (leading edge of click). 150 mm/s minimum. Average 3 beats.
Kato et al. (2012) compared both methods against fetal magnetocardiography in 135 and 84 fetuses respectively. Overestimation of the electrical PR was virtually identical: 14.6% for LV in/out and 14.7% for SVC/Ao. Andelfinger et al. also found no significant time difference between the two methods. Switching from MV/Ao to SVC/Ao as a rescue method does not introduce a clinically meaningful numerical offset.
The SVC a-wave reversal is driven by atrial contraction against venous return — not by LV diastolic filling. This is why E-A fusion does not affect it. At any heart rate, the SVC a-wave remains identifiable provided the angle and filter are correct.
Validated by Carvalho et al. (Royal Brompton, 2007). Less limited by fetal position but technically demanding — and values run systematically higher than MV/Ao.
Figure 7. Colour Doppler showing left and right pulmonary veins and aorta (Ao). GA 25w2d. Use colour Doppler to confirm vessel identity before placing the sample gate.
Figure 8. PA/PV Doppler trace. PR 114 ms. GA 25w2d, 150 mm/s. Note values run 6–15 ms higher than MV/Ao — method-specific reference ranges are mandatory.
Data from Anuwutnavin et al. Prenat Diagn 2018. Mean values in milliseconds. The offset widens progressively across gestation — this is clinically significant at any threshold near 150 ms.
| GA (weeks) | LV In/Out mean | PA/PV mean | Difference | Clinical relevance |
|---|---|---|---|---|
| 16–19 | 110.8 ms | 118.2 ms | +7.4 ms | 2nd trimester range |
| 20–24 | 115.9 ms | 122.0 ms | +6.1 ms | 2nd trimester range |
| 25–29 | 117.1 ms | 126.0 ms | +8.9 ms | Offset begins widening |
| 30–34 | 118.1 ms | 129.9 ms | +11.8 ms | 3rd trimester — clinically significant |
| 35–38 | 121.8 ms | 137.3 ms | +15.5 ms | Largest offset — highest risk of misclassification |
Applying MV/Ao reference ranges to PA/PV measurements is not valid at any gestational age. At 35–38 weeks a normal PA/PV value of 137 ms would appear to approach the 150 ms threshold when compared against MV/Ao ranges.
Not a first-line PR surveillance tool. Best used when rhythm analysis rather than PR number is the primary question.
Figure 9. Colour TDI apical view showing left ventricle (LV). Requires true apical orientation with cardiac long axis parallel to the beam. Angle of insonation below 30°.
Figure 10. TDI Doppler trace. a' wave marks atrial wall motion (atrial contraction surrogate). s' wave marks ventricular systole. PR 116 ms. GA 23w0d, 150 mm/s.
Atrial landmark: late diastolic atrial wall motion (A' wave). Ventricular landmark: isovolumic ventricular contraction spike (IVC spike). The interval mirrors the PR interval.
Lower bias vs ECG than MV/Ao in theory — but greater variability in practice. Supplementary role only for routine mPR surveillance.
Most measurement errors originate from wrong machine settings, not wrong probe position.
Figure 11. Non-negotiable system parameters for accurate acquisition.
Anuwutnavin et al. Prenat Diagn 2018. Mean (SD) in milliseconds by method and gestational age group.
All pulsed-wave Doppler methods overestimate the true electrical PR. Degree and reproducibility vary. Data from Bergman 2006 and Nii 2006.
Figure 12. Because PA/PV values run 6–15 ms higher than MV/Ao across gestation, switching methods mid-surveillance without accounting for this offset creates the illusion of PR prolongation where none exists.
| Method | Correlation vs ECG | Mean Bias | Reproducibility | Rating | Clinical Role |
|---|---|---|---|---|---|
| LV In/Out | r = 0.82 | +18.7 ms* | Good | Default first-line | |
| SVC/Ao | r = 0.85 | +14.7%† | Good | High FHR / rescue | |
| PA/PV | ~r = 0.82 | +6–15 ms‡ | Moderate | Position fallback | |
| TDI | R² = 0.15 | +8.0 ms | Advanced | Rhythm analysis |
*Bergman 2006, neonatal ECG validation (n=22). †Kato 2012, fetal magnetocardiography (n=84). ‡Gestational age dependent — 6–9 ms in 2nd trimester, 12–15 ms in 3rd trimester (Anuwutnavin 2018). Nii et al. Heart 2006 (n=196 fetal echos).
Figure 13. Situational routing for method selection — assess FHR and fetal position first, method follows.
Use LV In/Out (MV/Ao) as default.
Default methodSwitch to SVC/Ao. Reduce wall filter first.
High FHR rescueTry PA/PV. Apply method-specific reference ranges.
Position fallbackDocument method, FHR, and GA. Never compare values from different methods across serial studies without accounting for systematic differences.
AlwaysRoutine mPR measurement in all fetal echos is not recommended by any current guideline. Its role is specific.
| Indication | Status |
|---|---|
| Anti-SSA/Ro positive pregnancy Weekly from 16–26 weeks |
Standard |
| Fetal arrhythmia characterisation AV interval essential for rhythm diagnosis |
Standard |
| Complex arrhythmia — AV relationship SVC/Ao or TDI preferred |
Standard |
| Fetal growth restriction Studied — not clinically adopted |
Not adopted |
| Channelopathy family history No guideline recommendation |
Niche |
| Routine all-fetal screening | Not indicated |
Sweep speed 150 mm/s · Wall filter to minimum · Confirm fetal position · Note GA and FHR
Foot of A-wave onset to leading edge of aortic opening click. Average 3 beats minimum.
FHR above 160 bpm · E-A fusion · Lateral heart · Ambiguous A-wave onset
Method used · FHR · GA · Reference range applied · Beats averaged